Description
Screen-printed electrodes (SPEs) are electrochemical platforms manufactured by printing a wide variety of inks over different substrates allowing rapid on-site analysis with high reproducibility, sensitivity and accuracy.
Notice. Metrohm Dropsens uses cookies to provide necessary website functionality, improve your experience and analyze our traffic. By using our website, you agree to our Privacy Policy and our Cookie Policy.
Screen-printed electrodes (SPEs) are electrochemical platforms manufactured by printing a wide variety of inks over different substrates allowing rapid on-site analysis with high reproducibility, sensitivity and accuracy.
ADD TO QUOTATION
Number of WEs: 1 WE
Working electrode material: Calcium ionophore/carbon
Working electrode diameter: 0.40 cm
Working electrode position: In the middle of the strip.
Auxiliary electrode material: Carbon
Reference electrode material Silver
Contacts material: Silver
Substrate material: Ceramic
Substrate size: Length: 3.38 cm x Width: 1.02 cm
Substrate thickness: 0.05 cm
Main features: Single to use sensors that eliminate the electrode maintenance. Specially designed to work with microvolumes, this inexpensive miniaturized solution presents high reproducibility making them ideal for working in decentralized assays or to develop (bio)sensors in environmental, clinical or agri-food areas. The different modifications applied to the working electrode provide an enhanced active area and improve the electron-transfer properties.
Possible applications: Measurement of Ca2+ by open circuit potentiometry (OCP) in a range of concentration 10-4 to 0.1 M (from 4 to 4008 ppm).
Packaging: 50 units per pack
Storage: At room temperature, protected from light in a dry place.
Suitable connectors: CAST, I-CAST, CASTDIR, CAC, CAC4MM, DSC, DSC4MM, CAST1X8
Customization options: The design, geometry, substrate material, size, inks or modification of this reference can be customized under your specifications and produced in large quantities upon request. This modification can also be applied to 4WE and 8WE format SPEs. For more information please contact us at info.dropsens@metrohm.com